Can Poor Breath-Test Effort Understate Serum Alcohol? What 480 Emergency Patients Show

In 480 emergency patients, poor breath-test effort correlated less strongly with measured serum alcohol. Learn why no individual BAC correction follows.

Short answer: In a retrospective emergency-department study of 480 patients with paired nonzero alcohol results, staff-rated poor exhalation effort was associated with weaker correlation between breathalyzer-reported alcohol and clinical serum ethanol than staff-rated normal effort. The breath–serum correlation was 0.63 in the poor-effort group versus 0.92 in the normal-effort group. Average breathalyzer-reported concentrations were 0.19 g/dL in both groups, while average serum ethanol was 0.29 g/dL with poor effort and 0.23 g/dL with normal effort. These are group associations—not a validated correction, a diagnosis of deliberately weak blowing, or evidence that any particular driver’s forensic blood-alcohol concentration was underestimated.

The complete original primary study is Stellpflug and colleagues, Western Journal of Emergency Medicine (2025), DOI 10.5811/westjem.24998. Its limited scope matters: it compared an emergency-department breath instrument against serum ethanol obtained as much as 60 minutes apart. It did not conduct simultaneous forensic venous whole-blood headspace gas chromatography or measure driving performance.

What precise forensic question did the study answer?

The investigators asked whether a clinical staff member’s contemporaneous judgment of exhalation effort corresponded to how closely a patient’s breath alcohol result tracked a serum alcohol result obtained during the same visit. The practical question was agreement between two different kinds of measurement under ordinary emergency care, not whether a particular person was legally impaired.

A useful interpretation starts with three separate questions: What did the breath instrument report? What did the chemistry analyzer measure in separated serum? How far apart in time were the samples? Only after those are resolved can an examiner consider whether a difference is compatible with effort, sampling time, specimen matrix, analytical variation, or several influences together.

These boundaries also distinguish this paper from a generic introduction to breath alcohol measurement. The present study’s narrow contribution is the relationship between subjectively rated breath effort and breath-versus-serum discordance.

How were 480 emergency-department patients selected?

This was a retrospective electronic-health-record review spanning 49 months at one large tertiary-care hospital. Included patients had a breath ethanol result with an operator-entered effort designation of either “normal” or “poor,” plus a clinical serum ethanol result within 60 minutes. Both results had to be greater than zero. The researchers did not randomly assign someone to breathe poorly or normally, and they did not run a standardized forced-expiration experiment.

There were 245 patients labeled normal effort and 235 patients labeled poor effort. The person administering the breath test—usually an emergency technician or nurse—chose the category using clinical judgment. The publication does not describe an instrument-measured expiratory-flow threshold that independently established the designation.

Excluding zero-valued results limits how far conclusions can extend to sober persons, low readings, or borderline detectability. The single-hospital retrospective sample also limits assumptions about other clinical settings, police testing programs, instrument models, and specimens obtained under formal evidentiary protocols.

Which instrument and specimen did each result represent?

Breath tests used an Alco-Sensor FST. Clinical serum ethanol tests used an Abbott ARCHITECT c8000. The authors report both result types using a g/dL concentration convention. A number displayed as a breath-test ethanol or blood-equivalent figure must not be silently relabeled as a directly assayed physical blood sample. The article does not supply a validated person-specific breath-to-blood conversion.

Serum is the fluid obtained after whole blood clots and is separated for laboratory testing. It is not the same physical matrix as venous whole blood used in many forensic alcohol investigations. The study’s clinical serum analyzer is also not a demonstrated headspace GC-FID measurement of forensic whole blood. For the distinction between matrix and analytical method, see the site’s guide to headspace GC-FID blood alcohol testing.

Nor did this research determine ethanol glucuronide, phosphatidylethanol, an alcohol ingestion timeline, or any brain or behavioral biomarker. Its direct observations were paired reported alcohol results, elapsed time, and a staff-recorded judgment about effort.

What were the original study’s principal numerical results?

Original study cohort comparisons (Stellpflug et al., 2025). Values are as printed in the article’s clinical reporting convention, not an independent forensic blood analysis.
Measure All patients Normal effort Poor effort
Patients, n 480 245 235
Mean breathalyzer-reported alcohol, g/dL convention 0.19 0.19 0.19
Mean separately assayed serum ethanol, g/dL 0.26 0.23 0.29
Mean within-patient breath minus serum, g/dL convention −0.07 −0.04 −0.10
Pearson breath–serum correlation, r 0.75 0.92 0.63
Reported 95% CI for correlation 0.70–0.82 0.84–0.96 0.53–0.74

The mean difference in the normal-effort group was 0.04 g/dL in magnitude; in the poor-effort group it was 0.10 g/dL. Those are descriptive averages from different groups of patients. It would be unsound to subtract 0.06 g/dL from, or add 0.06 g/dL to, a defendant’s breath reading. The source does not validate either adjustment.

A stronger group correlation does not establish that individual pairs agree closely enough for legal substitution. A correlation describes how two series co-vary. An agreement analysis asks how different the methods’ individual values are, with appropriate clinically or legally relevant limits. The original paper offers group differences and correlations; it does not publish an externally validated correction formula, individual prediction intervals, or an error guarantee for a forensic instrument.

How far apart were breath tests and serum collections?

The study accepted a maximum gap of 60 minutes. Blood collection occurred before the breath test for some patients and after it for others. This is important because both biological ethanol concentration and the displayed instrument result may change during the interval. The gap is not a harmless clerical detail when the objective is individual agreement.

Original timing and test-order counts across the 480 included emergency encounters.
Recorded comparison condition Patient count Why it matters
Less than 15 minutes apart 237 Closer to contemporaneous sampling, but not necessarily simultaneous.
16–30 minutes apart 112 Biological concentration may have changed in the meantime.
31–60 minutes apart 131 More time for absorption or elimination and other intervening factors.
Serum blood collected before breath 184 Test order can alter the direction of apparent differences.
Serum blood collected after breath 288 The reverse sequence presents a different timing problem.
Tests performed concurrently 8 Only a very small subset was contemporaneous by the source’s classification.

These time-band counts total 480; the order counts also total 480. The paper did not provide a patient-level model that isolates subjective effort from time gap, specimen differences, changing alcohol kinetics, and other confounding variables. In a legal case, the actual timestamp of each procedure matters more than assuming a study-wide group difference applies to a different person’s chronology.

Does the lower correlation prove poor blowing caused an inaccurate result?

No. The study supports an association, but the effort label was not experimentally assigned or objectively calibrated. A patient classified as poor effort could differ from the normal-effort group in numerous ways. Notably, average serum ethanol was higher in the poor-effort group. The original investigators acknowledged that a person with higher alcohol exposure might have more difficulty coordinating a breath maneuver, but their study did not establish that hypothesis as the cause.

Correlations can differ because of distributions of measured values, variation in testing time, selection of patients, specimen effects and measurement noise. The retrospective record did not establish the relative contributions of all these factors. The observed difference in r is therefore a warning to review the quality and context of an individual test, rather than a causal demonstration that reduced expiration by itself generates a specific error.

The staff label also does not establish intentional noncooperation, deception, malingering, intoxication severity, physiological lung limitation, or a particular mouth-alcohol condition. Those are different propositions and need their own observations and properly validated methods.

Why are breath and serum numbers not directly interchangeable?

A breath measurement samples exhaled gas under instrument-specific conditions. A serum result measures ethanol in a processed blood-derived liquid using a laboratory chemistry system. Instruments can express or display results using conventions that resemble blood concentration units without turning a breath sample into a vial of blood. The study’s table is a comparison of instrument-reported values; it is not direct proof that the same physical ethanol concentration was independently measured in an identical matrix.

For forensic blood analysis, the questions include collection site, anticoagulant and preservative, specimen identity, timing, headspace method, calibration and analytical uncertainty. For clinical serum results, questions include the analytical platform, calibration and specimen fraction. For breath testing, relevant questions include the specific instrument, breath profile and sampling protocol. The same numeric label alone cannot erase these distinctions. The site’s specimen collection guide explains why physical matrices and collection metadata must remain attached to the result.

Nothing in this retrospective comparison establishes a universal serum-to-whole-blood factor, and nothing supplies a defendant-specific correction for a low or questionable breath result.

Does a mean difference of 0.10 g/dL measure a breath instrument’s error?

No. In the poor-effort group the source’s mean of breath minus serum was −0.10 g/dL, but that is the combined observed difference between two clinical measurement pathways under the study’s conditions. It is not an analytical bias estimated against traceable forensic whole-blood reference material, nor is it an error rate applicable to all breath analyzers.

A true analytical method evaluation would address reference material, calibration, replicate precision, limits of quantitation, method-specific uncertainty, specimen comparability and an appropriate definition of acceptable agreement. Those matters cannot be inferred from the correlation coefficient alone. Also, the study does not provide separate validated prediction limits for every serum concentration or every documented effort category.

Because the breath and serum tests were separated by up to an hour, assigning the entire observed average difference to the patient’s exhalation would attribute to one variable what the research design could not isolate.

Could the result establish BAC at the time of driving or injury?

Not by itself. There was no controlled drinking timeline, no universal absorption-stage determination, and no independent earlier incident-time whole-blood sample. A case-specific estimate for an earlier time requires its own validated and defensible assumptions about specimen, collection time, relevant drinking history, absorption and elimination, and uncertainty. This study provides no person-specific kinetic parameter and no rule authorizing a direct backward calculation from an emergency-department breath record.

It also did not measure standardized field sobriety tests, reaction-time tasks, driving simulators, crashes, observed injuries, or prospective behavioral outcomes. Thus an operator’s subjective poor-effort entry and a numerical ethanol difference cannot establish whether a person drove unsafely, possessed a particular mental state, or caused a particular injury.

Population associations can help identify questions that should be investigated. They do not replace the actual record, including pharmacokinetic timing and uncertainty when that is genuinely relevant.

How should defense and prosecution experts review a disputed breath reading?

The first step is to obtain the original readout rather than relying only on a transcribed chart number. Determine the instrument model and version, its approved use, the sequence of attempts and any objective sample-acceptance flags. The study’s checkbox “poor effort” is not a substitute for the actual instrumental breath-profile data, if such data were collected.

Next, identify who made the effort judgment and when, distinguish a recollection from a contemporaneous electronic entry, and check whether the operator used any defined criteria. Obtain precise collection timestamps, laboratory specimen type, requested tests, analyzer methodology, reported units, QC and any documented treatment or events between collections. Compare like with like; if the clinical result is serum, do not silently treat it as forensic whole-blood ethanol.

Finally, separate what can be established from what remains uncertain. A credible analysis can say that the 2025 emergency-department study found materially less breath–serum correspondence when staff rated effort poor. It cannot use that paper as proof of a precise individual correction, a failed statutory evidentiary test, or a person’s actual clinical or legal impairment.

What are the study’s most consequential limitations?

The retrospective design allows selection and confounding. The normal-versus-poor judgment was subjective, with no standardized flow metric reported. The two tests were often noncontemporaneous, and the broad inclusion window reached 60 minutes. Both tests had to be nonzero. The poor-effort cohort had a higher mean serum ethanol concentration, complicating causal comparison. The study came from one hospital using one named breath device and one named laboratory chemistry analyzer; extrapolation to different instruments or police protocols is not established.

An individual method-comparison claim would also require appropriate estimates of the distribution of paired differences across the concentration range and careful scrutiny of sample timing. A point estimate for group mean discrepancy cannot tell a laboratory whether a specific observed difference is within its validated measurement uncertainty or is sufficiently small for a particular clinical or forensic decision.

The distinction between “correlation” and “accuracy” is especially consequential. The research offers a useful clinical signal that effort documentation may matter. It does not supply Bland–Altman limits, an individual prediction guarantee, a validated decision threshold for the effort checkbox, or an externally confirmed correction applicable to a forensic case.

What is the scientifically defensible bottom line?

In 480 retrospective emergency-department encounters, operator-rated poor exhalation accompanied weaker breath–serum correlation and a larger average negative breath-minus-serum difference than operator-rated normal exhalation. The finding supports reviewing breath-test execution and paired clinical sampling information when discrepant results appear.

It does not prove that breath-test effort was the sole cause, that the reported clinical g/dL breath values are direct physical whole-blood measurements, that every poor-effort test is invalid, or that a particular person’s earlier BAC, impairment or case causation can be calculated. Those questions need evidence and validation not supplied by this original study.

Frequently asked questions

1. Does “poor effort” automatically make a breath alcohol test invalid?

No. The investigators found an association between a subjective poor-effort label and weaker breath–serum correspondence. They did not validate a universal invalidation rule for every instrument or evidentiary testing program.

2. Was the comparison against forensic whole-blood gas chromatography?

No. The reference comparison was hospital laboratory serum ethanol measured using an Abbott ARCHITECT c8000 system, not an identified forensic whole-blood headspace GC reference assay.

3. How many patients had poor and normal breath effort?

There were 235 poor-effort and 245 normal-effort patients, for 480 included encounters. Both recorded ethanol values had to be nonzero and within 60 minutes.

4. Can 0.10 g/dL be added to an individual person’s breath reading?

No. That is the poor-effort cohort’s mean breath-minus-serum difference in magnitude, not a validated correction or error bound for an individual.

5. Does a higher correlation mean the two methods agree exactly?

No. Correlation measures co-variation; individual agreement needs an explicit difference analysis, appropriately chosen limits and an identified clinical or forensic decision context.

6. Can this study identify when a driver last consumed alcohol?

No. It did not reconstruct participants’ drinking histories or validate earlier-time BAC estimates. Breath–serum differences may be affected by sampling order and changing concentration within the 60-minute window.

7. Can the findings prove intoxication caused a crash or injury?

No. The study measured no driving or crash event, standardized behavioral impairment, or case-specific causation. The results concern the interpretation of paired alcohol measurements in a clinical setting.